22
1100 Series FD Reference Manual
1
Installing the Fluorescence Detector
Installing the Detector
1
Install the LAN interface board in the detector (if required), see
2
Place the detector in the stack or on the bench in a horizontal position.
3
Ensure the line power switch at the front of the detector is OFF.
4
Connect the power cable to the power connector at the rear of the detector.
5
Connect the CAN cable to other Agilent 1100 Series modules.
6
If a Agilent ChemStation is the controller, connect either
•
the GPIB cable to the detector or
•
the LAN connection to the LAN interface board in the detector.
7
Connect the analog cable(s) (optional).
Preparations
Locate bench space
Provide power connections
Unpack the detector
Parts required
Detector
Power cord, for other cables see below and
ChemStation and/or Control Module G1323A/B.
Figure 5
Front View of Detector
Line power switch
with green light
Status indicator
green/yellow/red