1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
9
You can now select a stored spectra and evaluate/print them.
The maximum should be found at 397 nm
±
3 nm.
10
Load the method WLEMTEST. The FLD will change into the multi-excitation
mode and scan in the range of the expected maximum of 350 nm
±
20 nm.
11
Repeat steps
to
12
You can now select a stored spectra and evaluate/print them.
The maximum should be found at 350 nm
±
3 nm.
Figure 39
Evaluating Online Excitation Spectra
N O T E
The spectra on the control module is a real raw data spectrum. Therefore it looks not that
smooth as on the Agilent ChemStation (uses the spline algorithm).