134
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Simple Repairs
On the following pages repairs are described that can be carried out without
opening the main cover.
Table 23
Simple Repairs
Procedure
Typical Frequency
Notes
Flow cell exchange
If application requires a different flow cell type or if
defective.
Complete Assembly
A wavelength calibration check should
be performed after replacement.
If the flow cell is removed and inserted,
then a quick calibration check is
performed. If this fails, you must do a
wavelength recalibration, see
“Wavelength Verification and
Calibration"
Flow cell flushing
If flow cell is contaminated
Leak sensor drying
If leak has occurred.
Check for leaks.
Leak handling System
replacement
If broken or corroded.
Check for leaks.