1100 Series FD Reference Manual
8
Theory of Operation
Communication Settings for RS-232C Communication
The communication protocol used in this instrument supports only hardware
handshake (CTS/RTS).
Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the instrument must be
powered up again in order to store the values in the non-volatile memory.
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.
Table 66
Communication Settings for RS-232C Communication
Mode Select
1
2
3
4
5
6
7
8
RS-232C
0
1
Baud rate
Data
Bits
Parity
Table 67
Baud Rate Settings
Switches
Baud Rate
Switches
Baud Rate
3
4
5
3
4
5
0
0
0
9600
1
0
0
9600
0
0
1
1200
1
0
1
14400
0
1
0
2400
1
1
0
19200
0
1
1
4800
1
1
1
38400
Table 68
Data Bit Settings
Switch 6
Data Word Size
0
7 Bit Communication
1
8 Bit Communication
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......