Introduction to the Fluorescence Detector
7
1100 Series FD Reference Manual
247
The geometry of the grooves is optimized to reflect almost all of the incident
light, in the 1
st
order and disperse it with about 70% efficiency in the
ultra-violet range. Most of the remaining 30% of the light is reflected at zero
order, with no dispersion.
illustrates the light path at the surface of
the grating.
The grating is turned using a 3-phase brushless DC motor, the position of the
grating determining the wavelength or wavelength range of the light falling
onto the flow cell. The grating can be programmed to change its position and
therefore the wavelength during a run.
For spectra acquisition and multi-wavelength detection, the grating rotates at
4000 rpm.
The excitation and emission gratings are similar in design, but have different
blaze wavelengths. The excitation grating reflects most 1
st
order light in the
ultra-violet range around 250 nm, whereas the emission grating reflects better
in the visible range around 400 nm.
Figure 80
Dispersion of Light by a Grating
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......