34
1100 Series FD Reference Manual
2
First Steps with the Fluorescence Detector
Setting the Chromatographic Conditions
1
Set up the system with the following chromatographic conditions and wait
until the baseline gets stable.
Table 4
Chromatographic Conditions
Mobile phases
A = water = 35%
B = Acetonitrile = 65%
Column
OSD-Hypersil column, 125 mm x 4 mm i.d. with 5
µm particles
Sample
Isocratic standard sample, 1:10 diluted in
methanol
Flow rate
1.5 ml/min
Compressibility A (water)
46
Compressibility B (Acetonitrile)
115
Stroke A and B
auto
Stop time
4 minutes
Injection volume
5 µl
Oven temperature (1100)
30°C
FLD Excitations/Emission Wavelength
EX = 246 nm, EM = 317 nm
FLD PMT Gain
PMT = 10
FLD Response time
4 seconds