
Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
177
2
Remove the UV protector.
3
Locate the three slits (from left to right: PMT, EM, EX).
4
Use a pair of tweezers to replace the slit.
5
Put slight pressure on to the spring part of the slit.
Next Steps:
•
Replace the UV protector.
•
Replace the front foam.
•
Replace the covers as described in
“Replacing the Foams and Covers"
•
Re-install the flow cell and the front panel
•
Re-install the detector in the stack.
•
Turn the lamp on.
•
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......