
246
1100 Series FD Reference Manual
7
Introduction to the Fluorescence Detector
The radiation emitted by the lamp is dispersed and reflected by the excitation
monochromator grating onto the cell entrance slit.
The holographic concave grating is the main part of the monochromator,
dispersing and reflecting the incident light. The surface contains many minute
grooves, 1200 of them per millimeter. The grating carries a blaze to show
improved performance in the visible range.
Y
Figure 79
Mirror Assembly
Mirror
Grating (EX) inside
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......