178
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Removing the Optical Unit
When required
For following repairs: power supply, FLL board, Trigger Pack/FLF board and
PMT.
Tools required
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm
Parts required
Optical unit G1315-69002 (exchange assembly, comes with foam) or
individual parts depending on the following procedures
C A U T I O N
When working on the optical unit, a clean workbench with ESD protection mat must be
available. Otherwise optical components or electronic boards may be damaged.
N O T E
DO NOT remove the foam from the optical unit unless it is required during a procedure.
Otherwise parts can fall out.
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Disconnect capillaries.
•
Remove detector from stack and place it on the
workbench.
•
Remove the covers as described in
•
Remove the flow cell.
1
Disconnect cables (EM monochromator) and (EX
monochromator) from the detector main board (FLM).
A
B
A
B
A
EM
EX
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......