Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
185
6
Carefully pull off the ferret from the wires and keep it in a
safe place. Then pull out the FLL board.
7
Insert the new FLL board. Fit the wires through the holes
and the Ferret.
8
Put the wires into the released contacts using the
tweezers.
Order of wires (left to right):
9
Reinstall the lamp as described in
FLL Board
Ferret
FLL Board
Ferret
green
empty
brown
yellow
white
A
B