Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
165
Exchanging the Leak Sensor or Leak Panel
When required
If defective
Tools required
Screwdriver POZI 1 PT3
Flat screwdriver
Parts required
Leak sensor assembly 5061-3356
Leak Pan G1321-45501
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Remove the flow cell.
•
Remove detector from stack and place it on the
workbench.
•
Remove the covers as described in
•
Remove the flow cell.
1
Locate the leak sensor cable.