Preparation of the Flow Cell
1
Flush the flow cell with water.
2
Remove the inlet capillary from the flow cell.
3
Take the syringe and fix the needle to the syringe adapter.
4
Suck about 1.0 ml of the calibration sample into the syringe.
5
Keep the syringe in a horizontal position.
6
Remove t#he needle.
7
Add the filter to the syringe and fit the needle to filter.
8
Lift the needle tip and carefully eject approximately 0.5 ml to remove air
out of the syringe and to flush the needle.
9
Add the PEEK fitting to the needle tip and fix both at the flow cell inlet.
10
Slowly inject about 0.2 ml and wait for about 10 seconds to inject another
0.1 ml. This will assure that the cell is filled properly.
Figure 40
Syringe with Sample Filter
sample filter
N O T E
Do not inject the calibration sample without the sample filter.