
Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
97
Wavelength Calibration Failed
This message may show up during a wavelength calibration.
If the expected deviation is larger than the specified wavelength accuracy, the
message “Wavelength
Calibration Failed
” is displayed and the instrument
stays in a
Not Ready
condition.
Probable Causes
•
Flash lamp not ignited or position not correct.
•
Cell position not correct.
•
Solvent in the cell not clean or air bubble in the cell.
•
monochromator assembly position not correct (after replacement).
Suggested Actions
✔
Check the flash lamp image and position. See also
✔
Check the cell position.
✔
Flush the flow cell.
✔
Reset monochromator settings and re-run the wavelength calibration.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......