
136
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
2
Disconnect the capillaries from the flow cell.
3
Unscrew the thumb screws and pull the flow cell out of
the compartment.
Note:
The label attached to the flow cell provides
information on part number, cell volume and
maximum pressure. The cell type will be
automatically detected.
There are no parts that can be replaced on the flow
cell. If defective (leaky) the flow cell has to be
replaced completely.
4
Insert the flow cell and tighten the thumb screws.
Reconnect the capillaries to the flow cell. DO NOT install
the inlet capillary to the outlet connection of the flow cell.
This will result in poor performance or damage.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......