Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
149
5
Loosen the screws of the trigger pack holder a few turns.
6
Tip the holder to release the lamp and trigger pack. Slide
the cables of the trigger pack through the holder.
WARNING:
While pulling off the lamp be careful that you do not scrape your skin on the metal plates.
Do not remove the bottom pad. Otherwise the lamp will loose its correct position during reinstallation.