Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
155
Replacing the Lamp Cover Window
When required
If window is blocking the light throughput due to contamination
Tools required
Screwdriver POZI 1 PT3
Hexagonal screwdriver 2.5 mm, 100 mm long
Hexagonal screwdriver 4 mm, 100 mm long
Parts required
Quartz Window 1000-0999
N O T E
When loosening or removing screws take care that they do not fall into the instrument. This
may result in a complete removal of other assemblies.
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Remove the flow cell or disconnect the tubings from the
flow cell.
•
Remove detector from stack and place it on the
workbench.
•
Remove the covers as described in
1
Disconnect the emission and the excitation cables and
from the detector main board.
A
B
A
B
A
EM
EX