58
1100 Series FD Reference Manual
2
First Steps with the Fluorescence Detector
Evaluating The System Background
The example below uses water.
1
Pump solvent through your system.
2
Set the fluorescence scan range under FLD special setpoints according to
your needs.
3
Set PMT gain to 16.
4
Define a data file name and take a fluorescence scan. After the scan is
completed, the isoabsorbance scan results appear, see
N O T E
The scan time will increase when the range is enlarged. With the default values, the scan
takes about 2 minutes.
Figure 24
FLD special settings
The wavelength
range and step
number defines the
duration. Using the
maximum range, the
scan would take
approximately 10
minutes