Repairing the Fluorescence Detector
Replacing PMT and/or FLF board
When required
If defective or if application specific PMT is required.
Tools required
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm and 2.5 mm
Tweezers
Parts required
Photo-multiplier Tube (PMT) 1970-0201
Photo-multiplier Tube (PMT) for other application ranges, please contact
Hamamatsu dealers.
Alternative PMTs are either R928HA (185 to 900 nm) or R3788HA (185 to
750 nm). No other PMTs are recommended.
FLF Board G1321-69531 (requires firmware revision A.04.06 or above),
FLR Reference Diode board G1321-66533
C A U T I O N
When working on the optical unit, a clean workbench with ESD protection mat must be
available. Otherwise optical components or electronic boards may be damaged.
Preparations for this procedure:
•
Optical unit has been removed as described in
•
The foam must be in place. Otherwise cutoff filter and
diffuser will fall out.
1
Place the optical unit upside down on the ESD protection
mat.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......