First Steps with the Fluorescence Detector
2
1100 Series FD Reference Manual
43
Procedure II - Take two LC runs with the FLD
The conditions for the separation of organic compounds such as polyaromatic
nuclear hydrocarbons (PNAs) are well described in various standard methods,
including commonly used EPA and DIN methods. Achieving the best detection
levels requires checking for the optimum excitation and emission wavelengths
for all compounds. Yet taking fluorescence scans individually makes this a
tedious process. A better approach is to acquire spectra online for all
compounds during a run. This speeds up method development tremendously.
Two runs are sufficient for optimization.
During the
first run
, one wavelength is chosen in the low UV range for the
excitation wavelength and one emission wavelength in the spectral range for
the emission wavelength. Most fluorophores show strong absorption at these
wavelengths and the quantum yield is high. Excitation is sufficient for
collecting emission spectra.
Figure 13 on page 44 contains all emission spectra obtained in a single run
from a mix of 15 PNAs. This set of spectra is used to set up a timetable for
optimum emission wavelengths for all compounds.
The individual compound spectra in the isofluorescence plot show that at
least three emission wavelengths are needed to detect all 15 PNAs properly:
In the second run, three setpoints for emission wavelengths are entered into
the time-program and excitation spectra are recorded, as shown in figure 8.
The area of high intensity (red) is caused by stray light when emission spectra
overlap with the excitation wavelength. This can be avoided by fitting the
spectral range automatically. Excitation at 260 nm is most appropriate for all
PNAs.
Table 6
Timetable for PNA analysis
0 min:
350 nm
for naphthalene to phenanthrene
8.2 min:
420 nm
for anthracene to benzo(g,h,I)perylene
19.0 min:
500 nm
for indeno(1,2,3-cd)pyrene
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......