Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
127
Using the Built-in DAC Test
This function is available from the Control Module ONLY.
The DAC test provides a test pattern as an analog output. The output voltages
(analog 1 and analog 2) should show a constant value corresponding to the
ZERO OFFSET value. In addition to the constant voltage is a switched voltage
with a duration of 12 seconds and a height of 10 µV, see
1
Enable the function (
System - Tests - FLD - Enable DAC Test Pattern
).
2
Start the plot mode (e.g. Agilent 3396 with ATTN -4, ZERO=50, CS=5).
3
Stop the plot and disable the TAC Test (
System - Tests - FLD - Disable DAC
Test Pattern
).
4
Evaluate the noise (should be< 5 µV).
Figure 53
DAC Test (Example)
Limits:
Noise (pp) < 5 µV