Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
195
2
Reconnect the leak sensor cable (J3) using your left hand.
3
Reconnect the optical unit cable (J18) and the EM motor
cable (J20).
4
Insert the rear foam and reconnect cables (EM
monochromator) and (EX monochromator) to the detector
main board (FLM).
Next Steps:
•
Replace the covers as described in
•
Re-install the flow cell and the front panel
•
Re-install the detector in the stack.
•
Turn the lamp on.
•
Perform a wavelength verification to check the correct
positioning of the replaced assemblies, as described in
chapter
“Wavelength Verification and Calibration"
J20
J18
A
B
A
B
A
EM
EX
Rear
foam