Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
181
2
Unscrew screws A and B and disconnect connectors C
and D.
3
Disconnect the wires from the connector. Use a flat screw
driver to release the wires.
4
Remove the FLF board carefully in an upwards motion and
place it on an ESD protected mat.
5
If you need to replace either the board or the PMT
assembly, remove the PMT assembly from its socket.
6
Replace the PMT assembly or the FLF board.
7
Re-install the PMT assembly in the socket.
A
B
C
D
PMT assembly
Safety actuator
FLR board (REF
diode)