194
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Installing the Optical Unit
When required
When all repairs on the optical unit have been completed
Tools required
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm
WA R N I N G
For safe operation make sure that the ground cable is reconnected.
Preparations for this procedure:
•
All previous work has been completed.
•
Power supply has been installed.
•
Fan has been installed.
•
The optical unit has its foam cover attached.
1
Place the optical unit in the instrument. Reconnect the
grounding screw.
J3
Screw