OPERATION
LSM 510
Image Acquisition (Acquire)
B 40-051 e 07/98
4-29
To switch on the Enterprise laser, proceed as follows:
(1)
The internal water cooling LP 5 is running.
(2)
Start the PC, wait until NT system is booted.
(3)
Switch on the Power Supply of the Enterprise laser, power potentiometer turned to maximum.
(4)
Start the LSM 510 software.
If the LSM 510 software is already running and you want to use the UV laser, do the following:
(1)
Close the LSM 510 software.
(2)
Switch on the power supply of the Enterprise / power potentiometer turned to maximum.
(3)
Start the LSM 510 software again.
This applies to HeNe lasers:
After selecting the laser, click on the <On> button.
After a short heat-up phase of about 15 seconds, "Status ready - On" appears.
The required laser for image acquisition is now available.
Click on the <Close> button to close the "Laser Control" window.
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...