OPERATION
Image Acquisition (Acquire)
LSM 510
4-70
B 40-51 e 07/98
Click on the <Define Region> button.
The "Bleach Regions" window appears.
Define the required bleach regions in the scan
image.
Select the required laser wavelength and its
intensity under
Excitation of Bleach Track
in
the "Bleach Control" window.
The bleaching process will be started via the
<Bleach> button. However, it is also possible to
start the bleaching process via <Bleach> button
in the "Time Series Control" window and to
combine it with a time series.
The bleaching process can be finished via
<Stop> in the "Bleach Control" window.
☞
<Stop> does not only stop the
bleaching process, but the entire
scanning process.
Fig. 4-70
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...