![background image](http://html2.mh-extra.com/html/zeiss/lsm-510-inverted/lsm-510-inverted_operating-manual_3312162284.webp)
INTRODUCTION TO LASER SCANNING MICROSCOPY
Functions
LSM 510
5-30
B 40-051 e 07/98
Output
defines the name of the result sequence. It will get only those channels which are chosen by the
Input
parameter. The buttons labeled with 8 and 12 define the grey value (intensity) resolution in bit.
Normally the result will get the same resolution as the
Input
sequence. A change will be needed if image
sequences with different resolutions should be combined. Rising the grey value range to 12 bit will not
enhance the display quality or measurement accuracy. The smooth and morphology functions will
produce results with finer gradations.
The
Output
histogram shows the resulting histogram. They are not editable. The horizontal axis
represents the grey values from 0 to the maximum, which is either 255 or 4095, depending whether the
input is 8 bit or 12 bit. The vertical axis represents the pixel count. The vertical scale of the histogram is
set using the scroll bar. The units are percentages of the grey value distribution maximum. This setting
has no influence on the function.
Parameters:
Input
Input image sequence
Output
Output image sequence
Threshold
Exclusion value - 0...1000
Input L
Lower boundary of grey value range Input
Input C
Center of grey value range
Input
Input H
Upper boundary of grey value range
Input
Output L
Lower boundary of grey value range
Output
Output C
Center of grey value range
Output
Output H
Upper boundary of grey value range
Output
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...