OPERATION
LSM 510
Annex
B 40-051 e 07/98
4-191
To use the monitordiode for ratio application the following steps have to be done:
Activate the ratio channel R1 or R2 in <Ratio Settings> of the menu <Configuration Control> in
addition to the monitor diode channel (ChM-1) and one PMT channel.
Choose the appropriate PMT channel as source 1 in <Ratio settings> and ChM-1 as source 2. If this
numbering is changed (inverted) the ratio image will show a invertation of gray levels if compared to
the PMT image.
Fig. 4-178
☞
It is not possible to do the ratio between an on-line ratio image generated with two PMT
channels (as in ion-concentration sensitive ratio imaging) and the signal of the monitor diode.
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...