OPERATION
Annex
LSM 510
4-196
B 40-051 e 07/98
Default settings are 150 and 3000 respectively. With the help of the range indicator the default value is
changed until there is no pixel overflow anymore (‘red pixels’)
☞
Any new value can be set by hand typing and pressing the <ENTER> key while the scan is
running.
Any change in the setting parameters of PMT and diode signal will make a new Gain x in the
ratio formula necessary.
If the adjustment of all parameters is finished only the ratio image can be scanned respectively displayed
by switching off the PMT channel and the diode channel in the "Configuration Control" window and
leaving only the Ratio Channel be turned on. As a result only the ratio image is displayed; which can still
be influenced by the settings in PMT and diode channel.
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...