OPERATION
Image Optimization
LSM 510
4-88
B 40-051 e 07/98
On the
Channel Settings
panel of the "Scan Control" window, set the PMT gain with the
Detector
Gain
slider.
The image should not have more than a trace of red.
Adjust the
Ampl. Offset
slider so that areas without picture content just show a trace of blue.
If necessary, re-amplify brightness with the
Ampl. Gain
slider.
☞
Do not change the
Ampl. Gain
setting unless the settings made so far are insufficient for
optimizing the image.
Fig. 4-97
In the
Color Palette List
panel of the "Color Palette" window click on "No Palette".
This deselects the "Range Indicator" and activates the new presentation.
In the "Scan Control" window click on the <Stop> button.
This stops the continuous scan.
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...