OPERATION
LSM 510
Annex
B 40-051 e 07/98
4-195
Gain and offset in Ratio channel
If the setting of the PMT channel is finished, the range of the ratio channel is adjusted by the parameters
in the corresponding formalism. There are three types of formulas offered, when the button <R1-1> is
pushed. The only formula needed for a ratio image with the monitor-diode is type 1:
S1+n
S2+m
* x + y
The values for n and m have to be zero, as well as the value for y. Any deviation from zero will decrease
the contrast of the ratio image.
Only the value of x shall be influenced by the user. Dependent on the choice of data depth (8 or 12 bit),
x is between 0 and 256 (8bit) or between 0 and 4096 (12bit).
Fig. 4-181
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...