OPERATION
LSM 510
Image Acquisition (Acquire)
B 40-051 e 07/98
4-31
With
Transmitted Light
activated
, the halogen lamp is automatically occluded in the laser scanning
mode.
Please bear in mind that the light intensity does not automatically correspond to 0 % when
Light
Remote
is deactivated.
The microscope setting (light intensity) of the last session, which was not
remote-controlled, is restored on exit of the program.
In the
Objectives
list box, select the required objective as follows:
Open the list box.
Click on the objective you want to select.
The selected objective will automatically move into the beam path.
☞
Note that the nosepiece of the Axiovert 100 M microscope must not be moved manually. To
bring another objective into the beam path, move the objective nosepiece only software-
controlled.
Fig. 4-33
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...