OPERATION
Image Optimization
LSM 510
4-92
B 40-051 e 07/98
Click on the <Optimal Position> button in the
Collimator
box. Optimum positioning of the
collimator will be performed. The <Default
Position> button enables the collimator to be
set back to the factory-adjustment.
Note:
A change of the pinhole diameter made manually
in the
Pinhole
box will not be activated in the
Scan Control
box. Therefore, changes must
always be made in the
Channel Settings
box.
A filter change in
Autoadjust
is not displayed in
"Config. Control"
window.
Configuration 1 is equipped in such a way that
pinhole adjustment for channel 1 can only be
made with
= 488 nm, NFT 545, NFT 610 or
NFT 570.
Fig. 4-101
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...