OPERATION
LSM 510
Annex
B 40-051 e 07/98
4-185
4.12.6
Specification of Trigger-Interface LSM510
Application:
With the LSM510 Release 2.01 you can control various actions externally using Trigger-In or force
external devices to work at a defined time depending on an action using Trigger-Out during time series.
These actions are: Scan-Start/Stop, Bleach, Change of Scan-Interval, end of a countdown or even a
mouse-click on a button.
Interface:
Front plate Scanner-Interface (Scan-IF) inside
Elektronic-Box (Scan-Control-Module) of LSM510:
Connector ‘User I/O’, 26-pin shrinked SUB-D
Number:
4x Trigger-In, 4x Trigger-Out
Type/Voltage Range:
TTL (HCMOS), 0.0 - 5.0 V
Load:
In: 22 kOhm input impedance
Out:± 4 mA
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...