OPERATION
LSM 510
Analysis of Images and Stacks
B 40-051 e 07/98
4-107
(7) Select - Anim
This option allows you to call up the individual slice images of a stack in continuous succession
(animation).
The additional
Animate
dialog box allows you to influence the animation action.
Fig. 4-116
Using the left button unit, the stack can be passed automatically. The right button unit allows you to leaf
through the stack manually.
Using the <Speed1> and <Speed2> buttons, the preadjusted running speed can be interrogated.
With the <Increment> function, you can define the step width within the stack.
Only these step-by-step images will be shown in a running stack sequence.
(8) Select - Reuse
The parameters used for the creation of an image are stored together with the relevant image data. The
<Reuse> function allows the parameters of the loaded image to be automatically set in the instrument
and to continue working with this setting.
Control the
animation of
slices in a stack
Button unit
right
Button unit
left
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...