LSM 510 - SETUP REQUIREMENTS
LSM 510
Dimensions ... / Envirenment Requirements / Vibrations
B 40-051 e 07/98
2-7
2.4
Dimension of shipment crates
Crate containing
Length (cm)
Width (cm)
Height (cm)
Weight (kg)
Large system table
160
85
95
120
Small system table
90
75
80
80
LSM
190
85
120
350
Monitor, computer
120
80
90
80
UV laser unit
125
55
50
100
UV cooling unit
120
60
90
50
2.5 Environmental
Requirements
Operation, specified performance
T = 22 °C ±3 °C
Operation, reduced performance
T = 10 °C to 35 °C
Storage, less than 16h
T = -40 °C to 55 °C
Storage, less than 6h
T = -55 °C to 70 °C
Temperature gradient
±3 °C/h
Warm up time
1 h
Relative humidity
< 65 % at 30 °C
2.6 Vibrations
Vibrations under operation conditions
(with system table)
Shipping shock (LSM 5 box)
5 µm pp at 5 Hz
3 g
10 µm pp at 10 Hz
10 µm pp at 20 Hz
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...