11
1)
While the push rod to transmitted or reflected is to the working
position “R”, fluorescent microscopy can be used.
2)
Reduce the field diaphragm, If it is not centered, move it to the
center using the special tool. Adjustment can be done by opening the
field diaphragm.
3)
The push rod to direct image to binocular eyepieces or camera port
to the upper left of the trinocular head will send 100% of light to the
binocular eyepieces or to the top port. It satisfies high-grape
photography, see figure 7.
Preventative Maintenance
1. Cleaning frame and stage
Disconnect the plug from mains socket before cleaning. Clean the frame
and stage with a soft cloth moistened with a mild detergent solution. Be
sure the instrument is dry before using.
2. Cleaning optical parts
Microscope eyepieces and objectives are coated. They should not be
wiped while dry as dirt or dust may scratch the coating. It is best to
remove parts from the frame prior to cleaning. Always blow loose dust
away first. Use cotton swabs or lens tissue moistened with a lens cleaner
or a small amount of alcohol, then wipe the surface clean with a good
quality lens tissue. Solvents such as Xylene should NOT be used as
cleaner.
3. Cleaning 100X oil immersion lens
The immersion oil should be removed from the lens at the end of each
workday using cotton swabs or lens tissue moistened with a lens cleaner
or a small amount of alcohol.
LED Epifluorescence Microscope
L3201LED
User Guide
HAGAVISH 3,HOLON, ISRAEL
TEL: 972-3-5595252, FAX. 972-3-5594529