OPERATION
Image Optimization
LSM 510
4-98
B 40-051 e 07/98
4.7.4
Channel Shift Function
The
Channel Shift
function is used to produce a congruent image with relation to the pixels of the
various channels.
This pixel correction function is particularly important in UV applications.
(1)
Click on the <Process> and <Shift> buttons from the LSM 510 tool bar. The
Channel Shift
menu
appears on the screen.
(2)
Select the channels required for processing in the
Shift
box by clicking on the <Ch1> or <Ch3>
buttons. A tick
will appear in the button when the channels are activated.
(3)
Use the scrollbar or the < > and < > buttons to select the pixel shift in the horizontal and
vertical direction.
(4)
Click on the <Apply> button to activate the setting. If <
Preview> is activated, a preview of the
shift is shown in a separate image window.
Fig. 4-106
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...