4.12.7 Monitordiode
The monitordiode is placed in the excitation ray path of the LSM 510 behind the beam splitter combining
the visible and the uv ray path and in front of the main beam splitter. Therefore it allows for checking
the laser input in terms of power and noise. With the attenuation filter wheel in front of the diode it is
possible to attenuate the laser power reaching the diode. It is not possible to select one line out of a few
excitation wavelength to be detected by the diode.
To activate the diode as a detector do the following:
Click on the corresponding button in the "Configuration Control" window of the LSM 510 software.
Fig. 4-176
Choose either <Frame> or <Line> scan
Summary of Contents for LSM 510 Inverted
Page 1: ...I LSM 510 Laser Scanning Microscope Operating Manual...
Page 4: ...INTRODUCTION LSM 510 IV B 40 051 e 07 98...
Page 10: ...INTRODUCTION LSM 510 X B 40 051 e 07 98...
Page 12: ...NOTES ON DEVICE SAFETY Contents LSM 510 1 2 B 40 051 e 07 98...
Page 22: ...NOTES ON DEVICE SAFETY Warning and information labels LSM 510 1 12 B 40 051 e 07 98...
Page 24: ...LSM 510 SETUP REQUIREMENTS Contents LSM 510 2 2 B 40 051 e 07 98...
Page 34: ...LSM 510 SETUP REQUIREMENTS Laser Module VIS Laser Module UV LSM 510 2 12 B 40 051 e 07 98...
Page 36: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 48: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 3 2 B 40 051 e 07 98...
Page 231: ...OPERATION LSM 510 Annex B 40 051 e 07 98 4 173 Fig 4 171 Change over of the Scanning Module...
Page 256: ...INTRODUCTION TO LASER SCANNING MICROSCOPY Contents LSM 510 5 2 B 40 051 e 07 98...