Guide for Fluorescence Imaging

 

1.

 

Initial Set-up

 

~ Log in to NUCore and start a reservation.

 

~ Turn on Microscope Controller (A)
~ Turn on LED main power (B) and press control knob (C) to see a flashing 

sign. 

 

~ Wait the microscope initialization to complete and launch LASX 

software. (Computer has no password). 

~ Put slide on scope and set up image acquisition. 
 

2.

 

Mount sample on microscope

 

~ Place slide on stage ‘coverslip up’.

 

~ For transmitted light imaging:  

    Select BF button (D) and press TL-Shutter (F) on LED screen. Focus 

on sample. 

 

~      For fluorescent light imaging:  

                     Select FLUO (E), then appropriate filter cube from buttons on LED 

screen (DAP – Blue, L5 – Green, RHO – Red, Y5 – Far Red), open IL-
Shutter (F).

 

Focus on sample.

 

 

~  Find cell/region of interest. Close shutter.  

NOTE:

 To switch light path from oculars to camera gently pull out slider 

bar located in microscope head near oculars on the left side. 
 

3.

 

Troubleshooting

 

~ No fluorescent emission by eye?

 

-

 

Check the LED is ON. Control knob should show 

 

-

 

Check that light path is going to the oculars. The slider bar

 microscope head should be inside.  

~ No image displayed on a screen in LIVE mode? 

-

 

Check that light path is going to the camera. The slider bar

 microscope head should be taken out.  

~  Still having problems?

 

-

 

Contact BIF staff for assistance

.

 

 

 
 

Leica DM6B Widefield 

Quick Start Guide 

 