CHAPTER 3 - ADDITIONAL SOFTWARE TOOLS
ELYRA 7
ZEN Configuration Tool
ZEISS
03/2019 V_02
000000-2262-999
7
2
ZEN Configuration Tool
The configuration tool provides access to the database of a system. Be aware that wrong entries
can lead to a malfunction of the system. Always keep a backup of the original database. The
database of the system is stored in C:/ZEN/database. Always make a copy of the database before
you start to configure it.
Do not change any main settings or your system will not work anymore.
The following procedure describes how to define a new filter for external detectors like NDD or BiG.2.
Once defined within the configuration tool (Fig. 9), these filters will also be accessible via ZEN.
•
Open the tool; it will refer to the last database used which should be the systems database.
•
Make sure you have a copy of this database.
•
In case this is not the right database click
Browse
and choose the database to be configured.
•
Click
Start configuring
.
Fig. 9
Configuration Tool for LSM