8.3
Compatibility and Feature List of Components
System
HP Z240
HP Z840 single CPU
HP Z840 dual CPU
Supermicro Storage
CPU
Core i5 (4-core)
Xeon (4-core)
Xeon (8-core total)
Xeon (12-core total)
RAM max (with 16GB
modules)
64GB
(16GB default)
128GB
(16 GB default)
256GB
(64GB default)
256GB
(64GB default)
HDDs by default
4TB no RAID
2TB RAID 1
4TB RAID 10
36TB RAID 5
HDD options
2x 4TB HDDs
(need 5,25“ mounts)
2-4x 4TB HDDs
(need 5,25“ mounts)
2x 4TB HDDs
(need 5,25“ mounts)
none
Swap
128GB M.2 SSD
256GB PCIe SSD
on RAID 10; optional
512GB PCIe SSD
256GB SSD
Graphics
Quadro K620 2GB
Quadro K620 2GB;
option Quadro M4000
Quadro K2200 4GB;
option Quadro M4000
Quadro K2200 4GB;
option Quadro M4000
Network
1x 1GbE onboard;
option 1x 10GbE card
2x 1GbE onboard;
option 1x 10GbE card
2x 1GbE onboard;
option 1x 10GbE card
2x 10GbE and 1x 1GbE
onboard; option 1x
10GbE card
Remarks
ELYRA: either PCIe SSD
or 10GbE card
8.4
Networking of ZEISS Microscopy Computers
ZEISS microscope computers can be networked. The main purpose of such networking is storage of image
data, faster processing, workspace economy and service.
Networking is not intended for experimental download of drivers, patches, OS system upgrades or untested
third party software. This will very likely harm the system installation.
Networking also requires a certain work discipline, especially when handling file transfer and internet access.
Any files, which might contain harmful code should not be transferred to a system computer. It is not
recommended to connect the system computers to the internet without protection.
It is recommended to operate the ZEISS microscope computers in a kind of shielded LAN or e.g. via VPN, and
to have only one or two (e.g. storage) computers connected to the internet.
These computer can handle download of service packs or remote access for the actual system computers.
Some third party software has been approved to run together with our ZEN software. Arivis is one of our
approved partners. Ask your local ZEISS imaging specialist for the current software partners of ZEISS.