ELYRA 7
Left Tool Area and Hardware Control Tools
ZEISS
03/2019 V_02
000000-2262-999
61
Select the acquisition configuration from the
Parameters
list by highlighting it or search for it
by typing its name in the
Search
field. An
Info
menu pops up (Fig. 66). Press
Yes
to delete the
acquisition parameters,
No
to keep the
parameters. Close the menu by pressing the
cross
to abort the operation.
To save acquisition parameters press the
Save
button (
). This opens the
Save Active
Acquisition Parameters as
panel (Fig. 67). Type
in an appropriate name in the
Name
field and
press
Ok
to save the acquisition parameters in the
data base or press
Cancel
to abort the process.
The following Acquisition Parameters are stored as
a setting:
−
Time series: Cycles and Interval
−
Z-Stack: Range, Slices, Interval, Use of Piezo
−
Bleaching: Start and Repeat bleach settings,
Iterations, Stop at specific intensity, Different
scan speed, Trigger settings, laser
wavelength and power
−
Tile scan: Number of tiles, Overlap, Bi-
directional settings, Online stitching settings
(Settings for Bounding Grid and Convex Hull
are not saved as acquisition parameters)
−
All settings of the Acquisition Tool and
Channels Tool.
−
All other parameters are ignored as
acquisition parameters.
If the
New
button is pressed, a new
Image window
is opened in the
Image Container
into which the
new image will be displayed. The next acquired image is displayed in this image frame, when it is the
selected one. This function can be used to avoid that an old unsaved image (acquired by
Live
,
Continuous
or
Snap
) will be overwritten by a new image.
Fig. 66
Info panel
Fig. 67
Save Active Acquisition Parameters as
panel