Center Screen Area / Image Containers - Display and Image Analysis
ELYRA 7
394
000000-2262-999
03/2019 V_02
In order that a filter is applied to the data set, the corresponding box has to be checked. Checking the
box leads immediately to a re-calculation of the data set and all tables, plots and SMLM images (
HR
channels) will be updated automatically (Fig. 661).
All
SMLM-Filter
settings will be applied to images acquired during
Online Processing
.
All
SMLM-Filter
settings will be independently applied for each slice of the
Z Slice
.
Please be aware that the last filter settings during one imaging session will be kept for the next
image (acquired or loaded from daved data). If these do not fit, an empty image could be the case.
When experiencing images with too few or no localizations You are advised to check the filter
settings first.
Fig. 661
HR channel with no localization precision filter active (left)
and with filter set to min = 10 nm and max = 20 nm (right)