![background image](http://html1.mh-extra.com/html/zeiss/elyra-7/elyra-7_operating-manual_3332235373.webp)
ELYRA 7
Center Screen Area / Image Containers - Display and Image Analysis
ZEISS
03/2019 V_02
000000-2262-999
369
6.18.4
SMLM-Drift View Control Tab
SMLM-Drift
is used to perform a drift correction on the
image. You can select the type of correction you want to
perform by the
Correction Type
drop down menu,
which is only available in the
Show all
mode (Fig. 601).
You can select between
−
Fiducials
: drift correction will be performed on
fiucial markers in the image (Fig. 602)
−
Model-based
: drift correction will be performed
on sample structures (image self-alignments)
Fig. 602
SMLM-Drift view control tab of the SMLM view tab in Show all mode – Fiducials selected as Correction
type
Fig. 603
SMLM-Drift view control tab of the SMLM view tab in Show all mode – Model-based selected as
Correction type
Fig. 601
SMLM-Drift view control tab of
the SMLM view tab – Correction
type drop down menu