2-29
IM 760301-01E
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Faunctional Description
Split Display of the Waveform and Assignment of Waveforms «For procedures, see
section 6.6.»
The screen can be evenly divided and the waveforms can be assigned to the divided
windows. The screen can be divided into up to four windows. This function is useful
when there are many waveforms on the screen. You can select the method of
assignment from the choices below:
• Auto
The waveforms that are turned ON are assigned in order according to the element
number to the divided windows, voltage first and then current.
• Fixed
The waveforms are assigned in order by element number in the order voltage and
current to the divided windows regardless of whether or not the display is turned ON.
• User
The waveforms can be assigned arbitrarily to the divided windows regardless of
whether or not the display is turned ON.
Display Interpolation of the Waveform «For procedures, see section 6.7.»
The waveform display data can be connected linearly to display the waveform smoothly.
• Linear Interpolation
Linearly interpolates between two points.
• Interpolation OFF
No interpolation is performed. Only the data points are displayed.
Graticule «For procedures, see section 6.7.»
A grid or cross scale can be displayed on the screen. You can also select not to display
the grid or cross scale.
Displaying Scale Values «For procedures, see section 6.8.»
The upper and lower limits of the vertical axis and the values at the left and right ends of
the horizontal axis (time axis) of each waveform can be turned ON or OFF.
2.7 Waveform Display