Waveform Display
Note
•
To prevent erroneous operation caused by noise, the trigger function uses approximately 2%
hysteresis if the crest factor is set to 3. For example, if the trigger slope is set to , a trigger
occurs if the input signal passes through the trigger level on the rising edge after the input
signal level falls approximately 2% below the trigger level. The trigger function uses
approximately 4% hysteresis if the crest factor is set to 6.
•
The trigger function for waveform display does not work while integration is in progress or is
suspended. Therefore, the start position of the waveform display (the signal level at the left
end of the screen) may not be stable. In addition, the measurement period of the numeric
data and that of the waveform data may not be synchronized.