Power Measurement
• Display Resolution
The maximum display resolution of the integration value is 999999. When the
integrated value reaches 1000000 counts, the decimal point shifts automatically. For
example, if 0.001 mWh is added to 999.999 mWh, the display shows 1.00000 Wh.
• Display When Integration Overflow Occurs
If either of the conditions below is met, the integration is stopped, and the integration
time and integration value are held.
• The integration time reaches the maximum integration time (10000 hours).
• The integration value of WP, q, WS, or WQ reaches the maximum or minimum
integration value given below.
• Maximum/Minimum Displayed Integration Value
Active power (WP):
±
999999 MWh
Current (q):
±
999999 MAh
Apparent power (WS):
±
999999 MVAh
Reactive power (WQ):
±
999999 Mvarh
• Integration When MAX Hold Function Is Enabled
The integration value is determined and displayed by summing the value that is
measured at every display update rate, irrespective of the MAX hold function (section
5.5).
• Integration When the Measured Value Exceeds the Measurement Limit
If the instantaneous voltage or current of the sampled data exceeds 300% of the
measurement range when the crest factor is set to 3, the value is considered to be the
value corresponding to 300% for the purpose of integration. If the instantaneous
voltage or current of the sampled data exceeds 600% of the measurement range
when the crest factor is set to 6, the value is considered to be the value corresponding
to 600% for the purpose of integration.
• Integration When Current Input Is Small
Integration is performed assuming a current of zero when the current input with
respect to the measurement range is as indicated below.
• When the crest factor is set to 3
When Irms is less than or equal to 0.3% or Imn or Irmn is less than or equal to 2%
• When the crest factor is set to 6
When Irms is less than or equal to 0.6% or Imn or Irmn is less than or equal to 4%
• Using an External Signal to Control Integration (Option)
On models with the 20-channel D/A output option, you can use the remote control
feature to use an external signal to start, stop, and reset integration. For details about
the remote control feature, see section 3.4 in the
Extension Function User’s Manual
IM760301-51E
.
• Backup during Power Failures
• If a power failure occurs when integration is in progress, the integration result is
held. When the power recovers, the integration result calculated up to the point
when the power failure occurred is displayed with the integration operation
stopped.
• If the integration is reset after the power recovers, integration can be started once
again.
5.11 Integration