11-6
IM 760301-01E
Error in System Operation
Code
Message
Corrective Action
Reference Section
901
Failed to backup setup data.
The system has been initialized.
3.4
Maintenance service is required.
902
System RAM failure.
Maintenance service is required.
–
903
System ROM failure.
Maintenance service is required.
–
904
System RAM failure.
Maintenance service is required.
–
906
Fan stopped.
Power off immediately. Maintenance
–
service is required.
909
Illegal SUM value.
Maintenance service is required.
–
910
RAM read/write error.
Maintenance service is required.
–
911
Memory bus error.
Maintenance service is required.
–
912
Fatal error in Communication-driver.
Maintenance service is required.
–
914
Time out occurs in Communications.
–
–
915
EEPROM SUM error.
EEPROM may be damaged.
–
Maintenance service is required.
919
Module installation condition and setup
The system has been initialized.
–
parameters do not match.
Maintenance service is required.
921
System Failed to Draw Display.
Maintenance service is required.
–
922
Failed in communication with DSP.
Maintenance service is required.
–
923
Transmit data abnormality from DSP.
Maintenance service is required.
–
926
The USB device’s power consumption
–
12.9
exceeded the capacity of the USB hub.
Note
If the system is not operating correctly or if servicing is necessary, try initializing the WT3000.
The WT3000 may recover.