11-4
IM 760301-01E
Code
Message
Corrective Action
Reference Section
806
Cannot be set or executed in the
–
–
FFT measurement mode.
807
Cannot be set or executed
–
–*
in the Cycle by Cycle measurement mode.
808
Cannot change this parameter during
–
–
numerical display.
809
Cannot change this parameter during vector
–
–
display.
810
Cannot change this parameter during bar
–
–
graph display.
812
Cannot be set while storing/recalling of data
–
–
being performed.
813
Cannot be set while integration is running.
Reset Integration.
5.14
820
Cannot be set or executed in the Flicker
–
–*
measurement mode.
821
Attempt made to start flicker measurement
Wait for the initialization to finish and the WT3000
–*
while flicker initialization is not finished.
to become ready. Then, start the measurement.
822
Attempt made to start flicker measurement
Reset the flicker measurement, wait for the
–*
while flicker measurement is in progress or
initialization to finish, and start the measurement.
result of flicker judgement is displayed.
823
Cannot change during CAL.
Wait until CAL is completed.
10.1
824
Attempt made to initialize flicker
Reset the flicker measurement and then execute
–*
measurement while flicker measurement is
initialization.
in progress or resultof flicker judgement is
displayed.
825
Cannot change limit settings while flicker
Can be changed only when the measurement is
–*
measurement is in progress.
reset or while the judgement result is being
displayed.
826
Cannot be set or executed when flicker
Reset the flicker measurement.
–*
measurement is not reset.
827
Illegal math expression.
Input a correct computing equation.
5.4
828
Cannot start flicker measurement when all
Perform judgement or move the observation period –*
periods of dmax measurement are finished.
if necessary and measure again.
829
Cannot judge flicker measurement until all
Perform judgement after the dmax measurement of –*
periods of dmax measurement are finished.
all observation periods is finished.
830
Cannot move period while flicker
–
–*
measurement is not ready.
831
Processing now.
Retry setting or execution again.
–*
836
Cannot be set or executed while Cycle by
Reset the cycle-by-cycle measurement.
–*
Cycle measurement is not reset.
841
Attempted to start integration after
Reset Integration.
5.14
integration time has reached its preset value.
842
Attempted to start integration while
–
5.14
integration is in progress.
843
Measurement stopped due to overflow
Reset Integration.
5.14
during integration or due to a power failure.
844
Attempted to stop integration even though
–
5.14
integration was not in progress.
845
Attempted to start integration while
–
5.14
integration was in progress or integration
mode was not selected.
846
Attempted to start integration while
–
5.14
measurement of peak overflow was
in progress.
847
Attempted to start integration in continuous
–
5.11
integration mode when integration preset
time was set to “0”.
11.2 Error Messages and Corrective Actions