Measurement Conditions
Explanation
In the wiring system settings, you can select whether to set the measurement range or
synchronization source of input elements in the same wiring unit collectively or
independently.
• Turning ON/OFF the Independent Setting of Input Elements
For example, assume that the wiring system on a model with three input elements is
set as follows:
Input elements 1 to 3: Three-phase, four-wire system (3P4W). Input elements 1 to
3 are assigned to a single wiring unit
Σ
A.
• ON
The measurement range or synchronization source can be set independently for
each input element included in the wiring unit.
• OFF
The measurement range or synchronization source of input elements 1 to 3 is
set to the same setting. This is convenient in that the settings of the
measurement range and synchronization source of all input elements included
in the wiring unit can be done simultaneously when measuring a three-phase
device.
• Settings That Are Matched between Input Elements When the Independent
Setting of Input Elements Is Turned OFF
• Measurement range (including auto range ON/OFF)
• Voltage/Current mode
• Synchronization source setting
• Settings That Can Be Set Independently Even When the Independent Setting of
Input Elements Is Turned OFF
• Scaling value (VT ratio, CT ratio, and power coefficient)
• Input filter (line filter and frequency filter)
These settings can be set independently for each input element regardless of whether
the independent setting of input elements is turned ON/OFF.
• If a wiring unit (
Σ
A or
Σ
B) is specified and the independent setting of input
elements is changed from ON to OFF, the measurement range and
synchronization source settings of each input element are changed as follows:
• The measurement range is set to the maximum range on all input elements
included in the wiring unit. The external sensor input range has precedence over
the direct input range for the current range.
• The voltage/current mode and auto range settings are changed to match the
settings of the input element with the maximum measurement range. If multiple
input elements with maximum measurement range exist, the settings of the input
element with the smallest input element number takes precedence.
• The synchronization source setting is changed to match the setting of the input
element with the smallest input element number among the input elements
included in the wiring unit.
4.2 Selecting Independent Setting of Input Elements